LEOPARD可调移液器集中运用了基于操作与人体工程学的设计以及*的创新性材料。LEOPARDS系列移液器结构坚固,单手操作,整支灭菌,精度高,以及维护简单,可靠性强,还可自行进行校准。9个不同量程覆盖0.1μl-10000μl,能充分满足各种移液需求。
产品特点
性能指标
LEOPARD S系列整支高温灭菌移液器:
型号 | 检测点(ul) | 容量 允许误差(%) | 重复性误差(%) |
S-2 | 0.2 | ±12.00 | ≤6.00 |
1 | ±3.00 | ≤2.00 | |
2 | ±1.50 | ≤0.70 | |
S-10 | 1 | ±2.50 | ≤1.20 |
5 | ±1.50 | ≤0.6 0 | |
10 | ±1.00 | ≤0.40 | |
S-20 | 2 | ±3.00 | ≤1.50 |
10 | ±1.00 | ≤0.50 | |
20 | ±0.90 | ≤0.30 | |
S-50 | 5 | ±2.00 | ≤1.50 |
20 | ±1.00 | ≤0.60 | |
50 | ±0.60 | ≤0.30 | |
S-100 | 10 | ±1.70 | ≤0.50 |
50 | ±0.80 | ≤0.24 | |
100 | ±0.80 | ≤0.15 | |
S-200 | 20 | ±1.50 | ≤0.50 |
100 | ±0.80 | ≤0.25 | |
200 | ±0.60 | ≤0.15 | |
S-1000 | 100 | ±0.80 | ≤0.30 |
500 | ±0.60 | ≤0.20 | |
1000 | ±0.60 | ≤0.15 | |
S-5000 | 1000 | ±0.70 | ≤0.30 |
2500 | ±0.50 | ≤0.20 | |
5000 | ±0.50 | ≤0.15 | |
S-10000 | 1000 | ±1.00 | ≤0.50 |
5000 | ±0.60 | ≤0.20 | |
10000 | ±0.60 | ≤0.15 |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |