毛细管离心机

参考价面议
具体成交价以合同协议为准

联系方式:李享 查看联系方式

联系我们时请说明是仪器网上看到的信息,谢谢!


该厂商其他产品

我也要出现在这里

SH120-1毛细管离心机采用微电脑控制,自动平衡,有刷电机驱动,免维护,基本无碳粉污染。 该微量血液离心机采用高强度冷轧板制造,具有造型美观、体积小、重量轻、大容量、低噪音、高效率等特点。采用微电脑控制,自动平衡,可无刷电机驱动,免维护,无碳粉污染。

详细信息 在线询价

毛细管离心机

SH120-1微量血液离心机采用微电脑控制,自动平衡,刷电机驱动,免维护,无碳粉污染。  该微量血液离心机采用冷轧板制造,具有造型美观、体积小、重量轻、大容量、低噪音、高效率等特点。采用微电脑控制,自动平衡,无刷电机驱动,免维护,无碳粉污染。 

微量血液离心机技术参数:

    型    号: SH120-1
    规    格: 直径1.5×75mm×24支毛细管 
    技术指标: 转速范围:启动~12000转/分,无级调速
    容    量: 直径1.5×75mm×24支毛细管
    定时范围:0-120分
    zui大离心力:15300×g
    包装尺寸:35*34*36cm
    毛    重:10kg
    净    重: 9kg 

毛细管离心机 

毛细管离心机

 

SH120-1 blood centrifugeUser manual

1.     Product Features:

SH120 blood centrifugeismade of high strength plastic, with a beautiful shape, small size, light weight, large capacity, low noise, high efficiency and other characteristics. It isblood microcomputer controlledautomatic balance.

2.     Product Description:
SH120 can be used for precise determination of blood hematocrit value, small amount of blood. Take small amount of blood, large capacity, easy to take blood, centrifugation time is short, accurate detection, easy to use, efficient, and fully furnished.

3.     Main Technical Parameter

Power supply: 220V, 50Hz, 230W         Weight: 10kgs

Max. Speed: 12000r/min                Max. RCF: 15300*g

Max. Capacity: 24 places for¢1.5×75mm Timer: 0 – 30 mins

Dimension: 328*310*246mm (L*W*H)

4.     How to use:

First, open the lid, remove the cap and aluminum cover put the 24pcs of glass capillary tubes into the notch, (each capillary blood applied basically the same) and cover with aluminum cover cap fixed, power on the power switch, power indicator light, and then selecting the desired working time and speed (can be selected according to user need), close the lid, then centrifuges is now into operation status.

5.     Note

a.      Do not allow any kinds of liquid or debris into the working chamber; otherwise it will damage the motor.

b.     The instrument housing should be properly grounded to avoid the whole damp accidents.

 

1. c.      When the centrifuge is running, please do not open the lid, you must wait until it stops, then you can open the lid, remove the glass capillary tube.

2. d.     if there is any fault when using the instrument, please contact the manufacturer.

常州金坛良友仪器有限公司

|

手机:15851901332

联系人:李享

电话:86-0519-82533733

传真:86-0519-82533766

(联系我时,请说明是在仪器网上看到的,谢谢!)